Stage
Click a stage to move this equipment along the pipeline.
Description
Bench-top AC-DC high-voltage power supply. Converts 90–230 VAC mains input to a clean, linearly-adjustable DC output up to 30 kV at 0.5 mA. Used as the primary HV source for bench thrust testing and dielectric breakdown work. 10-turn potentiometer for output adjustment, lockable to prevent accidental changes during a run; digital readout shows display × 100 V. Internal over-current and short-circuit protection plus an external HV switch and main power switch.
Overview
- Brand
- Analog Technologies
- Model
- AHVAC30KVR5MABT
- Supplier
- analogtechnologies.com
- Weight
- 1192 g
Electrical
- Input voltage
- 90 – 230 V
- Max output voltage
- 30,000 V
- Max output current
- 0.5 mA
- Operating temp
- -20 – 55 °C
Reference
Features
High precision High efficiency (≥70 % at full load) High output voltage stability — short-term drift ≤0.05 %/min, long-term drift ≤0.05 %/h Linear modulation of output voltage Over-current protection Short-circuit protection (≤0.1 mA instantaneous short-circuit current) Digital display for output voltage (display × 100 V) 10-turn potentiometer with lock for stable setpoint
Applications
X-ray machines Spectral analysis Non-destructive inspection Semiconductor manufacturing equipment Particle accelerators Capillary electrophoresis Particle injection Physical vapour-phase deposition (PVD) Electrospinning / nanofibre preparation Glass / fabric coating DC reactive magnetron sputtering
Operating instructions
1. Connect the AC 90–230 V input through the rear IEC inlet, fuse 250 V / 2 A. 2. Confirm the 10-turn potentiometer is at 0 before flipping the high-voltage switch. 3. Turn on the main power switch, then enable the HV switch. 4. Slowly rotate the potentiometer clockwise while watching the digital display. Output voltage = display reading × 100 V. 5. When the target voltage is reached, rotate the potentiometer lock clockwise to fix the setpoint. 6. The HV output is on a 1 m thick brown lead; the ground terminal is on the left panel.
Safety notes
The supply MUST be reliably earthed before energising. Never touch the HV lead unless the supply is switched off AND the load + internal capacitors have fully discharged. Wait at least 5 minutes after switching off before any contact. Do not operate in a humid environment. The operator must not be connected to ground. Despite internal protection, avoid short circuits at the HV output. Use a properly insulated discharge stick to bleed residual charge before any inspection. Maintain insulation between the HV output and the surrounding environment, especially in vacuum / low-pressure setups.
Other specs
Input quiescent current: 140–160 mA (typ. 150 mA at I_OUT = 0) Input current at full load: 350–450 mA (typ. 400 mA at I_OUT = 0.5 mA) Input voltage regulation: ≤0.05 % across 90–230 V input Output modulation linearity: ≤0.1 % Output voltage temperature coefficient: ≤0.01 %/°C from −20 to 55 °C Output voltage range vs temperature: 0.99 V_OUT to 1.01 V_OUT across the operating range Load regulation: ≤0.05 %/mA from 0 to 0.5 mA Storage temperature: −20 to 80 °C MTBF: 1,000,000 h Output lead: 4.5 mm × 1.0 m thick brown HV cable Power cord: 6.5 mm × 1.8 m
Notes
Naming convention reference: A=Analog Technologies, HV=High Voltage Power Supply, AC=90–230 VAC input, 30KV=30 kV max output, R5=0.5 max output current, MA=mA current unit, BT=Bench Top.